Technical Specifications
Ranges Resolution Accuracy Remarks
  Specifications - Inputs
milivolt -150 to 150 mV 0.001 mV ± 0.01 % FS *** Rinput > 10 MΩ
  -500 to -150 mV 0.01 mV ± 0.02 % FS auto-ranging
  150 to 2450 mV 0.01 mV ± 0.02 % FS  
volt -10 to 45 V 0.0001 V ± 0.02 % FS Rinput > 1 MΩ
mA -5 to 24.5 mA 0.0001 mA ± 0.02 % FS Rinput < 120Ω
resistance 0 to 400 Ω 0.01 Ω ± 0.01 % FS Excitation current 0.85 mA
  400 to 2500 Ω 0.01 Ω ± 0.03 % FS auto-ranging
frequency* 0 to 600 Hz 0.01 Hz ± 0.04 Hz Rinput > 50KΩ
  600 to 1300 Hz 0.1 Hz ± 0.2 Hz Voltage DCmax = 30 V
  1300 to 5000 Hz 1 Hz ± 2 Hz AC Signal from 0.3 to 30 V
auto-ranging
counter* 0 to 108 -1 count 1 count The same remark as frequency Pulses Frequency < 3000 Hz
Pt-100 -200 to 850 °C / -328 to 1562 °F 0.01 °C / 0.01 °F ± 0.1 °C / ± 0.2 °F IEC-60751
Pt-1000 -200 to 400 °C / -328 to 752 °F 0.1 °C / 0.1 °F ± 0.1 °C / ± 0.2 °F IEC-60751
Cu-10 -200 to 260 °C / -328 to 500 °F 0.1 °C / 0.1 °F ± 2.0 °C / ± 4.0 °F Minco 16-9
Ni-100 -60 to 250 °C / -76 to 482 °F 0.1 °C / 0.1 °F ± 0.2 °C / ± 0.4 °F DIN-43760
probe** -200 to 850 °C / -328 to 1562 °F 0.01 °C / 0.01 °F ± 0.1 °C / ± 0.2 °F IEC-60751
TC-J -210 to 1200 °C / -346 to 2192 °F 0.1 °C / 0.1 °F ± 0.2 °C / ± 0.4 °F IEC-60584
TC-K -270 to -150 °C / -454 to -238 °F 0.1 °C / 0.1 °F ± 0.5 °C / ± 1.0 °F IEC-60584
  -150 to 1370 °C / -238 to 2498 °F 0.1 °C / 0.1 °F ± 0.2 °C / ± 0.4 °F  
TC-T -260 to -200 °C / -436 to -328 °F 0.1 °C / 0.1 °F ± 0.6 °C / ± 1.2 °F IEC-60584
  -200 to -75 °C / -328 to -103 °F 0.1 °C / 0.1 °F ± 0.4 °C / ± 0.8 °F  
  -75 to 400 °C / -103 to 752 °F 0.1 °C / 0.1 °F ± 0.2 °C / ± 0.4 °F  
TC-B 50 to 250 °C / 122 to 482 °F 0.1 °C / 0.1 °F ± 2.5 °C / ± 5.0 °F IEC-60584
  250 to 500 °C / 482 to 932 °F 0.1 °C / 0.1 °F ± 1.5 °C / ± 3.0 °F  
  500 to 1200 °C / 932 to 2192 °F 0.1 °C / 0.1 °F ± 1.0 °C / ± 2.0 °F  
  1200 to 1820 °C / 2192 to 3308 °F 0.1 °C / 0.1 °F ± 0.7 °C / ± 1.4 °F  
TC-R -50 to 300 °C / -58 to 572 °F 0.1 °C / 0.1 °F ± 1.0 °C / ± 2.0 °F IEC-60584
  300 to 1760 °C / 572 to 3200 °F 0.1 °C / 0.1 °F ± 0.7 °C / ± 1.4 °F  
TC-S -50 to 300 °C / -58 to 572 °F 0.1 °C / 0.1 °F ± 1.0 °C / ± 2.0 °F IEC-60584
  300 to 1760 °C / 572 to 3200 °F 0.1 °C / 0.1 °F ± 0.7 °C / ± 1.4 °F  
TC-E -270 to -150 °C / -454 to -238 °F 0.1 °C / 0.1 °F ± 0.3 °C / ± 0.6 °F IEC-60584
  -150 to 1000 °C / -238 to 1832 °F 0.1 °C / 0.1 °F ± 0.1 °C / ± 0.2 °F  
TC-N -260 to -200 °C / -436 to -328 °F 0.1 °C / 0.1 °F ± 1.0 °C / ± 2.0 °F IEC-60584
  -200 to -20 °C / -328 to -4 °F 0.1 °C / 0.1 °F ± 0.4 °C / ± 0.8 °F  
  -20 to 1300 °C / -4 to 2372 °F 0.1 °C / 0.1 °F ± 0.2 °C / ± 0.4 °F  
TC-L -200 to 900 °C / -328 to 1652 °F 0.1 °C / 0.1 °F ± 0.2 °C / ± 0.4 °F DIN-43710
TC-C 0 to 1500 °C / 32 to 2732 °F 0.1 °C / 0.1 °F ± 0.5 °C / ± 1.0 °F W5Re / W26Re
  1500 to 2320 °C / 2732 to 4208 °F 0.1 °C / 0.1 °F ± 0.7 °C / ± 1.4 °F  
(*) Accuracy valid since the frequency output is not configured.
(**) The Probe is a separate input used as reference thermometer. The related accuracy is relative only to the MCS-XV.
(***) FS = Full Scale.
  Specifications - Outputs
milivolt -10 to 110 mV 0.001 mV ± 0.02 % FS *** Rout < 0.3 Ω
volt -0.5 to 12 V 0.0001 V ± 0.02 % FS Rout < 0.3 Ω
mA 0 to 24 mA 0.0001 mA ± 0.02 % FS Rmax = 700 Ω
2-wire transmitter (XTR) 4 to 24 mA 0.0001 mA ± 0.02 % FS Vmax = 60 V
resistance 0 to 400 Ω 0.01 Ω ± 0.02 % FS For external excitation current
of 1.0 mA
  0 to 2500 Ω 0.1 Ω ± 0.03 % FS
frequency 0 to 100 Hz 0.01 Hz ± 0.02 Hz Peak value:
  0 to 10000 Hz 1 Hz ± 2 Hz 22 V / 25 mA max.
pulse 0 to 108 - 1 pulse 1 pulse Peak value:
22 V / 25 mA max.
Pulses frequency up to 10000 Hz
Pt-100 -200 to 850 °C / -328 to 1562 °F 0.01 °C / 0.01 °F ± 0.2 °C / ± 0.4 °F IEC-60751
Pt-1000 -200 to 400 °C / -328 to 752 °F 0.1 °C / 0.1 °F ± 0.1 °C / ± 0.2 °F IEC-60751
Cu-10 -200 to 260 °C / -328 to 500 °F 0.1 °C / 0.1 °F ± 2.0 °C / ± 4.0 °F Minco 16-9
Ni-100 -60 to 250 °C / -76 to 482 °F 0.1 °C / 0.1 °F ± 0.2 °C / ± 0.4 °F DIN-43760
TC-J -210 to 1200 °C / -346 to 2192 °F 0.1 °C / 0.1 °F ± 0.4 °C / ± 0.8 °F IEC-60584
TC-K -270 to -150 °C / -454 to -238 °F 0.1 °C / 0.1 °F ± 1.0 °C / ± 2.0 °F IEC-60584
  -150 to 1370 °C / -238 to 2498 °F 0.1 °C / 0.1 °F ± 0.4 °C / ± 0.8 °F  
TC-T -260 to -200 °C / -436 to -328 °F 0.1 °C / 0.1 °F ± 1.2 °C / ± 2.4 °F IEC-60584
  -200 to -75 °C / -328 to -103 °F 0.1 °C / 0.1 °F ± 0.8 °C / ± 1.6 °F  
  -75 to 400 °C / -103 to 752 °F 0.1 °C / 0.1 °F ± 0.4 °C / ± 0.8 °F  
TC-B 50 to 250 °C / 122 to 482 °F 0.1 °C / 0.1 °F ± 5.0 °C / ± 10.0 °F IEC-60584
  250 to 500 °C / 482 to 932 °F 0.1 °C / 0.1 °F ± 3.0 °C / ± 6.0 °F  
  500 to 1200 °C / 932 to 2192 °F 0.1 °C / 0.1 °F ± 2.0 °C / ± 4.0 °F  
  1200 to 1820 °C / 2192 to 3308 °F 0.1 °C / 0.1 °F ± 1.4 °C / ± 2.8 °F  
TC-R -50 to 300 °C / -58 to 572 °F 0.1 °C / 0.1 °F ± 2.0 °C / ± 4.0 °F IEC-60584
  300 to 1760 °C / 572 to 3200 °F 0.1 °C / 0.1 °F ± 1.4 °C / ± 2.8 °F  
TC-S -50 to 300 °C / -58 to 572 °F 0.1 °C / 0.1 °F ± 2.0 °C / ± 4.0 °F IEC-60584
  300 to 1760 °C / 572 to 3200 °F 0.1 °C / 0.1 °F ± 1.4 °C / ± 2.8 °F  
TC-E -270 to -150 °C / -454 to -238 °F 0.1 °C / 0.1 °F ± 0.6 °C / ± 1.2 °F IEC-60584
  -150 to 1000 °C / -238 to 1832 °F 0.1 °C / 0.1 °F ± 0.2 °C / ± 0.4 °F  
TC-N -260 to -200 °C / -436 to -328 °F 0.1 °C / 0.1 °F ± 2.0 °C / ± 4.0 °F IEC-60584
  -200 to -20 °C / -328 to -4 °F 0.1 °C / 0.1 °F ± 0.8 °C / ± 1.6 °F  
  -20 to 1300 °C / -4 to 2372 °F 0.1 °C / 0.1 °F ± 0.4 °C / ± 0.8 °F  
TC-L -200 to 900 °C / -328 to 1652 °F 0.1 °C / 0.1 °F ± 0.4 °C / ± 0.8 °F DIN-43710
TC-C 0 to 1500 °C / 32 to 2732 °F 0.1 °C / 0.1 °F ± 0.5 °C / ± 1.0 °F W5Re / W26Re
  1500 to 2320 °C / 2732 to 4208 °F 0.1 °C / 0.1 °F ± 0.7 °C / ± 1.4 °F  
The values of accuracy cover one year period and for a temperature range between 20 and 26 °C. Outside this range, the thermal stability is 0.001% FS / °C with reference
to 23 °C. Thermocouple with internal cold junction compensation, one must consider the error of this cold junction compensation of up to ± 0.2 °C or ± 0.4 °F.
  Specifications - Pressure
(0) 0 - 250 mmH20 0,001 ± 0,05 % FS Gage pressure.
Used with air or inert gases.
(1) 0 - 1 psi 0,0001 ± 0,05 % FS
(2) 0 - 5 psi 0,0001 ± 0,025 % FS
(3) 0 - 15 psi 0,0001 ± 0,025 % FS Gage or absolute pressure.
Used with fluids (gases or liquids) compatible with 316 L stainless steel.
(4) 0 - 30 psi 0,0001 ± 0,025 % FS
(5) 0 - 100 psi 0,001 ± 0,025 % FS
(6) 0 - 250 psi 0,001 ± 0,025 % FS
(7) 0 - 500 psi 0,01 ± 0,025 % FS
(8) 0 - 1000 psi 0,01 ± 0,025 % FS
(9) 0 - 3000 psi 0,01 ± 0,025 % FS
(10) 0 - 5000 psi 0,1 ± 0,025 % FS
(11) 0 - 10000 psi 0,1 ± 0,05 % FS
(12) Others on request